Update Time:2011-05-27   Writer:arima

National standards for the People's Republic
GB / T 2794-1995
Instead of GB2794-81
Determination of the viscosity adhesive
Determination methods for viscosity of adhesives
1 Subject Matter and Scope
This standard specifies the use of rotary viscometer and the viscosity of the viscosity method of Adhesives Cup.
The standard rotational viscometer for Newtonian or near Newtonian behavior of the adhesive viscosity.
The viscosity of the standard cup method for 50mL sample time 30 ~ 100s out of the adhesive viscosity.
2 Principle
2.1 rotational viscometer measured the viscosity is the dynamic viscosity, which is based on the apparent viscosity changes with shear rate change was reversible.
2.2 Viscosity Cup Viscosity measurement of viscosity is the condition, which is based on a volume of adhesive in a certain temperature, the diameter of the hole from the provisions of the outflow of time that the viscosity of the Song.
3 instruments and equipment
3.1 rotational viscometer.
3.2 The constant temperature bath: to maintain 23 ± 0.5 ℃ (adhesive can also be asked to choose other temperature).
3.3 Thermometer: Indexing is 0.1 ℃.
3.4 container: diameter of not less than 6cm, 11cm height of not less than the container or on a rotational viscometer with the container.
3.5 Viscosity Cup :1-4 viscosity cup capacity greater than 50mL. Specifications and dimensions shown below:
D holes are:
d (1) = 1.778 ± 0.003
d (2) = 2.54 ± 0.003
d (3) = 3.81 ± 0.003
d (4) = 6.35 ± 0.003
1-4 Viscosity Cup
3.6 Stopwatch: accuracy of 0.2s.
3.7 cylinder: 50mL.
3.8 Room temperature: to maintain 23 ± 0.5 ℃
4 specimen
4.1 The sample should be evenly without air bubbles.
4.2 The sample size to be able to meet the rotational viscometer and the viscosity needed to test the Cup.
5 Test procedure
5.1 rotational viscometer:
5.1.1 The sample should choose the same rotor and appropriate the same speed, so that the dial reading of 20% to 80% range.
5.1.2 will be filled with the sample container into the constant temperature bath, so that the sample temperature and test temperature balance, and maintain the sample temperature uniform.
5.1.3 vertical rotor immersed in the sample center of the rotor and to achieve the level surface marking (to be fitted with a protective frame.)
5.1.4 starting rotation viscometer, read pointer in the disc rotation constant readings.
5.1.5 Determination of three times for each sample.
5.2 Viscosity Cup:
5.2.1 wiped viscosity cup, and air dry or dry with a cold, light holes should observe the viscosity cup out clean.
5.2.2 cup on the sample and the viscosity constant room temperature.
5.2.3 Viscosity Cup and 50mL graduated cylinder to the vertical fixed on the stent, the outflow from the cylinder bottom hole 20cm, and out of holes in the viscosity cup placed below a 50mL graduated cylinder.
5.2.4 blocked out of hand holes, filling the viscosity of the sample cup.
5.2.5 Release your finger to the sample flow. Record your fingers out of holes to accept the removal of sample to 50mL graduated cylinder when the time to flow time (s) as the sample viscosity.
5.2.6 do it again the determination of the difference between the second measured value should not exceed the average of 5%.
6 The results indicated that
6.1 rotary viscometer method, three samples taken a reading test, the minimum value, whichever is the effective number three.
6.2 Viscosity Cup, the results of the arithmetic mean value, whichever is the effective number three.
7 determination results
7.1 rotational viscometer: the readings are calculated according to the provisions viscometer to Pa? S or mPa? S said.
7.2 Viscosity cup: to s said.
8 Test report
Test report should include the following:
a. Sample source, name, type;
b. Rotary viscometer model used in the rotor, the speed or the viscosity cup with numbers;
c. Test temperature;
d. Viscosity.
Additional Information:
The standard proposed by China Ministry of Chemical Industry.
This standard is under the jurisdiction of the Shanghai Rubber Products Institute.
This standard is responsible for drafting the Shanghai Rubber Products Institute.
Main drafters of this standard Fen Ren Xin.
This standard is equivalent to the standard American Society for Testing and Materials ASTM D 1084-1988 "Determination of viscosity adhesives."